Testing SIMIND with a Heart Phantom

Monte CarloSat Nov 23 2024
Advertisement
Simulating medical images with nuclear detectors, known as Monte Carlo simulation codes like SIMIND, are really handy tools for scientists. This study was all about checking if SIMIND could be a good substitute for real nuclear medicine experiments. They used a cardiac phantom, called HL, which is like a fake heart, to compare SIMIND's results with actual experiment data. They performed single-photon emission computed tomography/computed tomography (SPECT/CT) imaging using a Discovery NM/CT 670 scanner on this phantom. This imaging was done with a special type of iodine, ¹²³I-meta-iodobenzylguanidine, that's used in myocardial scintigraphy. During the simulation, they looked at three energy windows: the main one (159 keV ± 10%), a lower one (137. 5 keV ± 4%), and a higher one (180. 5 keV ± 3%). Both the real experiments and simulations were done under the same conditions.
They picked some important areas, called regions of interest (ROIs), in the phantom and the simulated data. They then calculated the mean, maximum, and minimum counts in each of these areas. They also made a polar map for the heart parts and figured out the relative errors for each segment. The results were mostly good. The relative errors were less than 10% for most data from the main energy window after reconstruction. However, in the lower and higher energy windows, the errors were sometimes more than 10%. The smallest errors happened when looking at the average counts in the ROIs. In the polar map, the errors were higher at the base of the heart but stayed below 10% from the middle to the tip. So, while SIMIND can be a good alternative for some nuclear medicine experiments, it might not always be super accurate, especially in certain situations. It's like having a really good backup plan, but you still need to be careful and check things twice.
https://localnews.ai/article/testing-simind-with-a-heart-phantom-86568eb4

actions